In the process and manufacturing industry, pressure is one of the variables that most often need to be recorded and monitored. Whether in standard applications or under special operating conditions, each application requires pressure gauges that are perfect for their needs. In Germany, pressure gauges are calibrated by accredited calibration laboratories, according to guideline DKD-R 6-1 (DKD-R 6-1, 2014). This calibration guideline establishes minimum requirements for the calibration procedure and the measurement uncertainty calculation when calibrating pressure gauges. In addition to the uncertainty contributions due to the calibration, the uncertainty contributions due to the specific application, like extreme temperatures, high pressure in containers, extreme height differences, shocks, aggressive media or problematic physical product properties, should be taken into account. This paper presents the approach in which the measurement uncertainty can be calculated in industrial pressure measurements. Furthermore, the individual uncertainty contributions and their identification or origin are discussed. Finally, an example of a measurement uncertainty budget is shown as an important tool in the measurement uncertainty calculation.
In industry, various types of pressure gauges like bourdon tube manometers,
electrical pressure gauges, or pressure transmitters with electrical output
are used. These types differ in their construction, functionality, and
measuring principle. Pressure gauges are available in various designs: more
than 150 measuring ranges are available from approximately 25 mbar to 15 000 bar
in the most important internationally used units as gauge pressure, absolute
pressure, and vacuum, but also as
All of the pressure gauges can be calibrated by accredited calibration laboratories according to guideline DKD-R 6-1, “Calibration of Pressure Gauges”. This guideline was drawn up by the “Pressure and Vacuum” DKD Technical Committee, in cooperation with the PTB and the accredited calibration laboratories. It specifies minimum requirements for the calibration procedure and for the measurement uncertainty in the calibration of pressure gauges. It takes into account the three different types of pressure:
positive and negative gauge pressure; absolute pressure; and differential pressure.
The calibration of pressure gauges is performed by direct comparison of the
measured values of the calibration object with those of the reference
standard. The metrological traceability of the measuring equipment is
ensured by the accredited calibration laboratories.
Depending on the desired measurement uncertainty, pressure gauges are calibrated according to different sequences and their respective measurement series number. The different calibration sequences are shown in Table 1.
Calibration sequences according to DKD-R 6-1.
For the measuring setup, it is important to be certain that the calibration object and the standard are at the same height (reference height), to avoid pressure differences due to height differences. The calibration of pressure gauges is carried out at equally distributed measuring points across the calibration range which correspond to the range of application of the pressure gauges. Depending on the desired measurement uncertainty or accuracy of the calibration object, the measurements are carried out at different numbers of pre-loadings and measuring series (see Table 1). The pre-loadings serve the function of wetting the calibration setup with the pressure medium, and for improving the elasticity of the measuring systems. The magnitude of measurement uncertainty is directly related to the number of measurement series.
During calibration, each pressure stage is first approached in the upward measuring series. After setting the pressure and keeping a waiting time of at least 30 s, the calibration object and the standard are compared with each other. This waiting time is necessary to stabilize the pressure throughout the measurement system and to indicate the applied pressure on the calibration object and on the standard (if digital). This happens only after a certain relaxation time. Especially for Bourdon tube manometers, the frictional effect of the pointer system is minimized by a slight tapping on the housing of the manometers. For the pressure step variation in a measurement series, the time between two successive load steps is the same.
After reaching the highest measuring value of the calibration object, a waiting time of 2 min (for Bourdon tube manometers 5 min) is observed before the measurement is continued with the downward measuring series.
When approaching the individual load steps, care is taken that the measuring points are not exceeded in the upward measuring series or have not fallen below in the downward measuring series.
The determination of the measurement uncertainty is done for each calibration measurement value, meaning for every load step.
The calibration of pressure gauges is to be carried out at stable ambient
temperature in the range of 18 to 28
Sufficient temperature equilibrium between the calibration object and the environment, as well as a warm-up time of the calibration object (or a possible warming of the calibration object by the supply voltage), are considered.
In addition to the uncertainty contributions due to the calibration, the uncertainty contributions due to the specific application, like extreme temperatures, high pressure in containers, extreme height differences, shocks, aggressive media or problematic physical product properties should be taken into account. The user of the pressure gauge should do additional investigations to quantify these uncertainty contributions which are relevant for the specific application.
The measurement uncertainty is defined in VIM (ISO/IEC-Leitfaden 99, 2007) as a non-negative parameter characterizing the dispersion of the values being attributed to the measure, and is based on the information used.
Terms and calculation rules for measurement uncertainty determination according to the guide to the expression of uncertainty in measurement (JCGM 100, 2008) and EA guideline “Evaluation of the Uncertainty of Measurement In Calibration” (EA-4/02 M, 2013) are listed in Table 2. Correlations between the input quantities are not taken into account.
Terms and calculation rules for measurement uncertainty determination.
Source: DKD-R 6-1 calibration of pressure gauges (DKD-R 6-1, 2014).
The following Fig. 1 shows an example of the possible influences for determination of expected value and measurement uncertainty, which should be considered when calibrating a pressure gauge with a piston gauge (pressure balance) as a standard. Based on this illustration, the model of evaluation and calculation of the measurement uncertainty are prepared (Heidenblut et al., 2007; Sommer et al., 2006).
Possible influence quantities for calibration of pressure gauges.
In addition to the uncertainty contributions due to the calibration process, application influences should be considered. For example, the temperature gradients could be greater than during calibration. In this case, additional investigations of the pressure gauge should be done to determine the magnitude of this influence.
For calibration of Bourdon tube manometers and digital pressure gauges, a
simple sum/difference model for determining the mean measurement deviation
of the indication is used:
In Table 3, all used variables are listed and explained.
Overview of the used variables.
When using the averages from the up and down measurement series, the
expanded uncertainty (
When measuring pressure gauges, piston gauges (pressure balances), electrical pressure gauges, or pressure calibrators are used as a reference standard.
When using piston gauges, it should be considered that the expanded
calibration uncertainty which is given in the calibration certificate is
valid for reference conditions (standard acceleration or local acceleration,
at a temperature of 20
For a piston gauge under application conditions, two corrections must be applied. The first correction takes into account the influence quantities such as
which in turn are associated with individual measurement uncertainties.
The temperature of the piston is measured with a calibrated thermometer during the current calibration. The measurement uncertainty of the thermometer is taken from the calibration certificate.
The thermal expansion coefficients and the coefficient of deformation of the piston–cylinder unit, including the associated measurement uncertainties, are given in the calibration certificate of the piston gauge.
The acceleration of gravity is known with sufficient accuracy for the location of the measurement. The uncertainty contribution of the acceleration of gravity is not significant and can be neglected. The uncertainty contribution of the deformation coefficient depends on pressure. At pressures above about 100 MPa, the deformation coefficient becomes a dominant uncertainty source.
Table 4 shows the measurement uncertainty budget for the temperature influence on the calibration procedure.
Measurement uncertainty budget due to temperature influence on the measurement procedure.
Another correction takes into account the influence quantities for determination of hydrostatic pressure:
In turn, these influence quantities are assigned individual measurement uncertainties.
The difference in altitude between the pressure reference levels of the pressure gauge and of the piston gauge must be currently determined for each measurement. The measuring equipment used for this purpose is a calibrated height gauge or a steel ruler. It should be noted that the pressure reference level of the piston gauge corresponds to the lower edge of the piston in measuring position.
The density difference is calculated from the current air density and the density of the pressure-transmitting medium. The current air density is determined from the ambient temperature and the atmospheric pressure, which means taking into account the pressure-dependent gas density (approximation).
The density of the pressure-transmitting medium is given in the calibration certificate of the piston gauge. The acceleration of gravity at the measuring location of the measurement is considered. Table 5 shows the measurement uncertainty budget for the influence of the difference in altitude on the measurement procedure.
Measurement uncertainty budget due to difference in altitude to the measurement procedure.
When using directly indicating pressure gauges (digital pressure gauges/calibrators) as standard measuring instruments, the following influence quantities are taken into account in the measurement uncertainty budget.
Measurement uncertainty (calibration certificate, normal distribution) Resolution (manufacturer specification, rectangular distribution) Long-term stability (drift between the recalibrations, rectangular
distribution) If necessary, the temperature dependence (manufacturer specification,
rectangular distribution)
Source: DKD-R 6-1 calibration of pressure gauges (DKD-R 6-1, 2014)
The resolution of the indicating device is obtained from the ratio of the
pointer width to the centre distance of two adjacent graduation lines (scale
interval). In practice, the ratio
The best estimate of an analogue indicating device is determined by visual
interpolation. The smallest estimable fraction of a scale interval is the
interpolation component
The resolution of digital indicating devices corresponds to the digital
step. For the determination of the uncertainty contribution, half the value
of the resolution
If the reading fluctuates by more than the previously determined value of
the resolution with the pressure gauge not being loaded, the resolution
The zero point (unloaded pressure gauge usually at atmospheric pressure) is
set prior to each measurement cycle and is recorded before and after each
measurement cycle. The reading is carried out with the instrument being
completely relieved. The zero deviation is calculated according to the
following equation:
The repeatability with the mounting not being changed is determined from the
difference of the zero signal-corrected measurement values of the corresponding
measurement series.
When stating the mean values, the hysteresis is determined from the
difference of the zero point-corrected measurement values of the increasing
and decreasing series as follows:
Possible influence quantities due to the specific application become necessary if the boundary conditions used by the manufacturer for the calculation of the total performance are different from the site of application of the pressure-measuring device (Industrielle Druck-Messtechnik Grundlagen und Praxis, 2017).
Each measurement result of measuring instruments is influenced by changing environmental and process conditions and by the spatial arrangement of the measuring instruments. For two reasons it is therefore important to carry out an estimation of the expected measuring uncertainty of a measurement setup:
ensuring that a selected measurement setup can capture the required process
parameters with the accuracy required by the process; and selection of the pressure gauge from a number of different types, which is
best suited for a measurement task with regard to the expected measuring
uncertainty.
To determine the expected measuring uncertainty, it is first necessary to
know the relevant environmental and process conditions. The following
questions should be answered.
In which range does the ambient temperature change? In which range does the process temperature change? In which range does the process pressure change? Is the transmitter exposed to vibrations at the application site? Is the power supply constant or is it fluctuating? Do electromagnetic influences affect the pressure gauge? Is there a specific alignment of the pressure gauge at the application site? How are the electrical connections realized? Are there any electrical
disturbing quantities?
The influences and effects of changes in the ambient and process conditions on the pressure gauge are specified in the manufacturer's specifications. If the application of the pressure gauge differs from these conditions, additional investigations should be planned to quantify an additional uncertainty contribution.
In this paper, the influences of the ambient or process temperature, vibrations in the pipeline and non-constant power supply are taken into account.
Additional investigations should be done if the ambient or process temperature of the specific application differs from the boundary conditions of the manufacturer. This influence can be quantified by repeated measurements at different simulated temperatures. To determine the measurement uncertainty contribution, the half-width of the rectangular distribution is assigned half of the span of the measured values.
Typical values are
Additional investigations should be done if there are strong vibrations in
the measuring system, which can influence the pressure gauge. This influence
can be quantified by monitoring the pressure in the measuring system
generated by the pressure control unit with a reference pressure gauge at
different pressure points. To determine the measurement uncertainty
contribution, the half-width of the rectangular distribution is assigned
half of the span of the measured values. Typical values are <
Separate investigations should be done if the power supply is not constant
or it is fluctuating. To determine the measurement uncertainty
contribution, the half-width of the rectangular distribution is assigned
half of the span of the measured values. Typical values are less than
These uncertainty contributions can be combined into the uncertainty contribution for the specific application.
Sensitivity coefficients
The input/influence quantities are preferably summarized in a measurement uncertainty budget (Krystek, 2012). Table 6 below shows a general representation of a measurement uncertainty budget. In Table 7, an example measurement uncertainty budget in measurement of a digital pressure gauge is shown.
Measurement uncertainty budget in calibration of digital pressure gauges.
Example measurement uncertainty budget in measurement of digital pressure gauges, sequence A, measuring range 0 to 5 bar, pressure medium nitrogen.
As a result of the uncertainty analysis or the measurement uncertainty budget, the complete measurement result of the digital pressure gauge is composed of the determined deviation of 0.0014 bar and the associated expanded measurement uncertainty of 0.019 bar (taking into account a coverage probability of approximately 95 %).
According to guideline DKD-R 6-1 when calibrating pressure gauges,
regardless of the result of the calibration, the measurement uncertainty is
specified
for calibration according to sequence B not smaller than 0.04 % of
the measurement span and for calibration according to sequence C not smaller than 0.30 % of
the measurement span.
These limiting values are taken into account when specifying the measurement
uncertainties.
Application of smart sensor technologies and digitally networked measurement systems creates new challenges in terms of calibration and traceability. In industry, the number of sensors is increasing rapidly.
The challenge is to deal with this multitude of sensors that cannot be removed from the system, so that classical calibration is not possible.
In addition, correlation is a very important issue in this area and needs to be taken into account as it can significantly change uncertainty. Furthermore, the sensors may have a different or dynamic behaviour.
Therefore, new calibration and traceability strategies and calibration of these sensors and measurement systems are required.
A novel approach to solutions at lower hierarchy levels to ensure traceability is artificial intelligence, e.g. machine learning. Machine learning uses algorithms to analyse data, recognize patterns, learn from them, and then make a determination or prediction about something. Machine learning algorithms are capable of solving many every day or even very specific problems.
In practice, problems often arise when there are either too few data or when there are too many dimensions of the data.
Instead of manually coding software routines with a particular set of instructions to perform a particular task, the machine is “trained” with large amounts of data and algorithms so that it can learn how to perform the task.
There are various kinds of pressure gauges available for industrial
pressure measurement. The calibration of these pressure gauges is carried
out according to guideline DKD-R 6-1. This guideline specifies requirements
for the calibration procedure as well as the calculation of the measurement
uncertainty. For the measurement uncertainty determination, measurement
uncertainty contributions of the used standard, the calibration procedure,
the ambient conditions, the calibration object, and the examiner are
included. These factors should be considered in the industrial application
of pressure gauges as well, meaning it is important to know the
repeatability or the hysteresis of a pressure gauge and to apply it in the
practical measuring process. Additional uncertainty contributions due to the
specific application of the industrial pressure measurement should be
investigated and quantified. For these influences a sub-budget
Furthermore, when using piston gauges, the necessary influence quantities, such as temperature dependence of the piston–cylinder unit or the difference in altitude, have to be taken into account or be corrected.
Finally, an example measurement uncertainty budget is presented, which includes all investigated uncertainty contributions and reflects their influence on the expanded measurement uncertainty during a measurement.
In addition, the new challenges in terms of calibration and traceability when applying smart sensor technologies and digitally networked measurement systems were addressed. Machine learning is an industrial trend of these years and needs to be further developed.
The underlying measurement data are not publicly available and can be requested from the authors if required.
NS performed the uncertainty calculations and wrote the article. OSW advised, reviewed and recommended the corrections of the article.
The authors declare that they have no conflict of interest.