Regular research article
14 Dec 2017
Electron capture detector based on a non-radioactive electron source: operating parameters vs. analytical performance
Erik Bunert, Ansgar T. Kirk, Jens Oermann, and Stefan Zimmermann Institute of Electrical Engineering and Measurement Technology, Department of Sensors and Measurement Technology, Leibniz Universität Hannover, 30167 Hannover, Germany
Abstract. Gas chromatographs with electron capture detectors are widely used for the analysis of electron affine substances such as pesticides or chlorofluorocarbons. With detection limits in the low pptv range, electron capture detectors are the most sensitive detectors available for such compounds. Based on their operating principle, they require free electrons at atmospheric pressure, which are usually generated by a β decay. However, the use of radioactive materials leads to regulatory restrictions regarding purchase, operation, and disposal. Here, we present a novel electron capture detector based on a non-radioactive electron source that shows similar detection limits compared to radioactive detectors but that is not subject to these limitations and offers further advantages such as adjustable electron densities and energies. In this work we show first experimental results using 1,1,2-trichloroethane and sevoflurane, and investigate the effect of several operating parameters on the analytical performance of this new non-radioactive electron capture detector (ECD).

Citation: Bunert, E., Kirk, A. T., Oermann, J., and Zimmermann, S.: Electron capture detector based on a non-radioactive electron source: operating parameters vs. analytical performance, J. Sens. Sens. Syst., 6, 381-387, https://doi.org/10.5194/jsss-6-381-2017, 2017.

Publications Copernicus
Special issue
Download
Short summary
Electron capture detectors (ECDs) are widely used for the detection of electron affine substances such as pesticides or chlorofluorocarbons. In this work we present a new non-radioactive ECD and investigate the analytical performance depending on the operating parameters. We achieved limits of detection for 1,1,2-trichloroethane and sevoflurane, which are comparable to radioactive ECDs. Furthermore, a pulsed collector voltage mode was implemented, leading to an extended linear range.
Electron capture detectors (ECDs) are widely used for the detection of electron affine...